EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>5977B Series

Agilent Technologies 5977B Series Troubleshooting And Maintenance Guide

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
General Maintenance 4
5977B Series MSD Troubleshooting and Maintenance Manual 143
To Adjust the Quad Frequency
Materials needed
Screwdriver, flat-blade, large (8730-0002)
Procedure
1 Ensure the MSD is at thermal equilibrium. It takes at least 2 hours after all
heated zones have reached their setpoints for the MSD to reach thermal
equilibrium.
2 Open the analyzer cover.
3 Ensure the RF cover on the side board is secure and no screws are missing.
A loose RF cover or missing screw can significantly affect coil adjustment.
4 In the Tune and Vacuum Control view, select Optimize Quadrupole Frequency from
the Execute menu.
5 Enter an m/z value of 100.
6 Slowly turn the quad frequency adjustment screws to minimize the voltage
displayed. (See Figure 22 on page 144.)
Turn the adjustment screws alternately. Turn each screw only a little bit at
a time. Keep the screws at equal extension.
7 When the voltage is minimized, click Stop.
WARNING
Do not remove any other covers. Dangerous voltages are present under these
covers.
CAUTION
Do not use a coin to adjust the screws. If you drop it, it could fall into the electronics
fan and cause significant damage.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals