214 5977B Series MSD Troubleshooting and Maintenance Manual
8 Electronics
Side Board
The side board is mounted on the side plate. It performs these functions:
Provides the 1 MHz reference clock for the RF amplifier.
Generates the RF component of the voltage applied to the quadrupole mass
filter according to a signal from the main board. The amplitude of this
voltage is proportional to the mass selected.
Generates the DC component of the voltage applied to the quadrupole mass
filter. The magnitude of this voltage is proportional to the RF voltage.
Passes voltages generated on the main board and the detector focus voltage
from the HED power supply to elements in the ion source and the detector.
Generates and adjusts filament emission current and electron energy as
controlled by the main board.
Switches the filament power from one filament to the other.
Monitors for RF faults and shuts down the analyzer if one is detected.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals