3 CI Troubleshooting
Common CI-Specific Problems
Because of the added complexity of the parts required for CI, there are many
potential problems added. The greatest number of, and most serious problems
with CI are associated with leaks or contamination in the reagent gas
introduction system. NCI is especially sensitive to the presence of air; small
leaks that cause no problems in PCI can destroy NCI sensitivity.
As with EI, if the MSD tunes well and no air leak is present, sample sensitivity
problems should be addressed by GC inlet maintenance first.
Wrong reagent gas
Reagent gas not hooked up or hooked up to wrong reagent gas inlet port
Wrong ions entered in tune file
Wrong tune file selected
Not enough bakeout time has elapsed since vent (background is too high)
Wrong column positioning (extending > 2 mm past tip of interface)
Interface tip seal not installed
EI ion source installed in CI mode
EI filament or other EI ion source parts in CI ion source
Air leaks in reagent gas flow path
CI filament has stretched and sagged:
High EMV
Linear (no inflection point) electron energy (EIEnrgy) ramp

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals