General Troubleshooting 2
5977B Series MSD Troubleshooting and Maintenance Manual 61
Contamination
Contamination is usually identified by excessive background in the mass
spectra. It can come from the GC or from the MSD. The source of the
contamination can sometimes be determined by identifying the contaminants.
Some contaminants are much more likely to originate in the GC. Others are
more likely to originate in the MSD.
Contamination originating in the GC typically comes from one of these
sources:
Column or septum bleed
Dirty GC inlet
GC inlet liner
Contaminated syringe
Poor quality carrier gas
Dirty carrier gas tubing
Fingerprints (improper handling of clean parts)
Contamination originating in the MSD typically comes from one of the
following sources:
Air leak
Cleaning solvents and materials
Foreline pump oil (standard foreline pump)
Fingerprints (improper handling of clean parts)
Table 6 on page 62 lists some of the more common contaminants, the ions
characteristic of those contaminants, and the likely sources of those
contaminants.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals