EasyManuals Logo

Agilent Technologies 5977B Series Troubleshooting And Maintenance Guide

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
130 5977B Series MSD Troubleshooting and Maintenance Manual
4 General Maintenance
To Purge the Calibration Valves
EI calibration valve
After adding new PFTBA to the calibrant vial, you must purge the air out of
the vial and valve.
1 If the vacuum gauge controller is on, turn it off.
2 In Tune and Vacuum Control view, select Purge Calibrant Valve under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
CI calibration valve
After adding new PFDTD to the calibrant vial, you must purge the air out of
the vial and valve.
1 If the vacuum gauge controller is on, turn it off.
2 Verify that PCICH4.U is loaded.
3 In Tune and Vacuum Control view, select Purge Calibrant Valve under the
Vacuum menu.
This will open the CI calibration valve for several minutes with all analyzer
voltages turned off.
CAUTION
After removing a calibrant vial, you must purge the calibration valve. Failure to do so
will result in damage to the filaments and the electron multiplier.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals