EasyManua.ls Logo

Agilent Technologies 5977B Series

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Maintenance 4
5977B Series MSD Troubleshooting and Maintenance Manual 135
To Remove the Ion Vacuum Gauge
Procedure
1 Vent the MSD.
2 Disconnect the cable on the back of the ion vacuum gauge.
3 Unscrew the red plastic thumbnut on the gauge clamp.
4 Remove the long screw from the clamp.
5 While supporting the gauge body, remove the clamp from the mounting
flange.
6 Remove the gauge.
7 If you will not be replacing the gauge soon, install the blanking plate
provided with the gauge and secure it with the clamp, screw, and
thumbnut.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Related product manuals