General Troubleshooting 2
5977B Series MSD Troubleshooting and Maintenance Manual 37
Troubleshooting Tips and Tricks
Rule 1: Look for what has been changed.
Many problems are introduced accidentally by human actions. Every time any
system is disturbed, there is a chance of introducing a new problem.
If the MSD was just pumped down after maintenance, suspect air leaks or
incorrect assembly.
If carrier gas or helium gas purifier were just changed, suspect leaks or
contaminated or incorrect gas.
If the GC column was just replaced, suspect air leaks or contaminated or
bleeding column.
Rule 2: If complex is not working, go back to simple.
A complex task is not only more difficult to perform but also more difficult to
troubleshoot. If you are having trouble detecting your sample, verify that
autotune is successful.
Rule 3: Divide and conquer.
This technique is known as half-split troubleshooting. If you can isolate the
problem to only part of the system, it is much easier to locate.
To determine whether an air leak is in the GC or the MSD, you can vent the
MSD, remove the column, and install the blank interface ferrule. If the leak
goes away, it was in the GC.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals