120 5977B Series MSD Troubleshooting and Maintenance Manual
4 General Maintenance
To Replace the DP Fluid
Materials needed
Aluminum foil, clean
Cloths, clean, lint-free (05980-60051)
Container for old DP fluid
DP fluid, 18.5 mL (6040-0809) – Two bottles are required
Gloves
Oil- and solvent-resistant
Thermally insulated
Procedure
1 Remove the DP from the MSD. (See “To Remove the DP” on page 118.)
Remove the O-ring assembly from the top of the DP.
2 Cover the top of the DP with aluminum foil (shiny side up).
3 Heat the DP at 60 °C for 15 minutes. (For 7820A or 7890 GC’s you can use
the GC oven.)
4 Pour the old DP fluid out the top of the pump. Even after heating, the
pump fluid pours very slowly.
5 Check the color of the pump fluid.
WARNING
The pump and pump fluid will be hot. Wear protective gloves when you remove the
pump from the oven.
WARNING
Treat the old pump fluid as hazardous. It may contain traces of toxic chemicals.
WARNING
Methylene chloride is a hazardous solvent. Work in a fume hood and take all
appropriate precautions.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals