EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>5977B Series

Agilent Technologies 5977B Series Troubleshooting And Maintenance Guide

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
60 5977B Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
Air Leaks
Air leaks are a problem for any instrument that requires a vacuum to operate.
Leaks are generally caused by vacuum seals that are damaged or not fastened
correctly. Symptoms of leaks include:
Higher than normal analyzer chamber pressure or foreline pressure
Higher than normal background
Peaks characteristic of air (m/z 18, 28, 32, and 44 or m/z 14 and 16)
Poor sensitivity
Low relative abundance of m/z 502 (this varies with the tune program used)
Leaks can occur in either the GC or the MSD. The most likely point for an air
leak is a seal you recently opened.
In the GC, most leaks occur in:
GC inlet septum
GC inlet column nut
Broken or cracked capillary column
Leaks can occur in many more places in the MSD:
GC/MSD interface column nut
Side plate O-ring (all the way around)
Vent valve O-ring
Calibration valve(s)
GC/MSD interface O-ring (where the interface attaches to the analyzer
chamber)
Front and rear end plate O-rings
Turbo pump O-ring

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals