200 5977B Series MSD Troubleshooting and Maintenance Manual
7 Analyzer
Filament care
Similar to the filaments in incandescent light bulbs, the ion source filaments
will eventually burn out. Certain practices reduce the chance of early failure:
If you have an optional G3397B Micro-Ion vacuum gauge, use it to verify
that the system has an adequate vacuum before turning on the analyzer,
especially after any maintenance was performed.
If you are controlling your MSD from the Manual Tune screen, always select
MSOff before changing any of the filament parameters.
When setting up data acquisition parameters, set the solvent delay so that
the analyzer will not turn on while the solvent peak is eluting.
When the software prompts Override solvent delay? at the beginning of a run,
always select NO.
Higher emission current reduces filament life.
Higher electron energy reduces filament life.
Leaving the filament on for short times (1 minute) during data acquisition
reduces filament life.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals