44 5977B Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
Poor sensitivity
Incorrect tuning
Tune file that does not match the type of analysis
Repeller voltage is too low
Incorrect temperatures (oven, GC/MSD interface, ion source, or mass filter)
Incorrect sample concentration
Leaking GC inlet
Dirty GC inlet
Incorrect split ratio
Purge off time in splitless mode is too short
Excessive pressure in the MSD
Dirty ion source
Air leak
Poor filament operation
Detector (HED electron multiplier) is not working correctly
Incorrect mass filter polarity
† This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
Dirty syringe needle
Dirty GC inlet
Leaking GC inlet
Injection is too large
Loose column connections
Variations in pressure, column flow, and temperature
Dirty ion source
Loose connections in the analyzer
Ground loops
† This could cause a fault condition in the GC that would prevent the GC
from operating.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals