General Troubleshooting 2
5977B Series MSD Troubleshooting and Maintenance Manual 53
Mass filter temperature sensor failed
Source power cable is not connected to the sideboard
MSD electronics are not working correctly
† These will cause an error message.
GC/MSD interface will not heat up
Incorrect setpoint(s)
Setpoint entered in wrong heated zone
GC/MSD interface has not had enough time to reach temperature setpoint
GC is off
GC experienced a fault and needs to be reset
GC/MSD interface heater/sensor cable is not connected
GC/MSD heater failed (burned out)
GC/MSD sensor failed
GC electronics are not working correctly
† These will cause a GC error message. GC error messages are described in
the documentation supplied with your GC.