58 5977B Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
The system is in standby
This message is triggered by a shutdown signal on the remote start cable. It is
usually caused by a GC fault, an ALS fault, or a bad cable connection. Once the
cause of the fault is corrected, selecting MS ON or checking MSD status should
remove the message.
The system is in vent state
Wait at least 30 minutes after seeing this message before you actually vent the
MSD.
System was vented on purpose (no problem)
Fan fault has turned off the high vacuum pump (power cycle the MSD to
clear the fault)
Fuse for the high vacuum pump has failed
MSD electronics are not working correctly
There is no emission current
Filament is not connected properly; try the other filament
Filament has failed; try the other filament
MSD electronics are not working correctly
There is not enough signal to begin tune
Corrupted tune file
Poor mass axis calibration
Amu gain or offset is too high
Calibration vial(s) empty or almost empty
Excessive pressure in the analyzer chamber
Air leak
CAUTION
Venting the MSD too soon after this message appears can damage the turbo pump.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals