EasyManua.ls Logo

Agilent Technologies 5977B Series

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76
3 CI Troubleshooting
No reagent gas peaks in NCI
Reagent gases do not ionize in NCI; look for background ions instead.
Verify tune parameters.
If no background ions are visible, go back to methane PCI.
No PFDTD calibrant peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Verify tune parameters.
Go back to methane PCI.
No sample peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Go back to methane PCI.
Poor quality reagent gas (purity less than 99.99%).
Large peak at m/z 238 in NCI OFN spectrum
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Find and fix your small air leak.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Related product manuals