3 CI Troubleshooting
No reagent gas peaks in NCI
Reagent gases do not ionize in NCI; look for background ions instead.
Verify tune parameters.
If no background ions are visible, go back to methane PCI.
No PFDTD calibrant peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Verify tune parameters.
Go back to methane PCI.
No sample peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Go back to methane PCI.
Poor quality reagent gas (purity less than 99.99%).
Large peak at m/z 238 in NCI OFN spectrum
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
).
Find and fix your small air leak.