EasyManuals Logo

Agilent Technologies 5977B Series Troubleshooting And Maintenance Guide

Agilent Technologies 5977B Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
CI Troubleshooting 3
81
Large peak at m/z 19
If the abundance of the peak at m/z 19 is more than half abundance of the
peak at m/z 17, there probably is too much water in the system.
The system was not baked out sufficiently after it was last vented.
Bake out the system as described in the Maintenance chapter of this manual.
Moisture left over in the reagent gas supply tubing and flow module
Purge the reagent gas supply lines for at least 60 minutes.
Contaminated reagent gas supply
Replace the reagent gas supply, and purge the lines and flow module.
Saturated methane/isobutane gas purifier
Replace the gas purifier.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals