Inlets
Agilent 6890 Gas Chromatograph Service Manual
Solvent Vapor Exit Accessory 260
Leak testing
Leak testing
When checking the solvent vapor exit assembly for leaks, first check the
tri-column assembly. If the tri-column assembly is not leaking, check the
valve/fitting assembly (which includes the stainless steel union, tubing, and
fitting on the solenoid).
Potential leak points
• Tri-column assembly
• Quartz column connector
• Quartz Y splitter
• Connection to the inlet
• Valve/fitting assembly
• Solenoid valve
• S/S union
• Inlet fitting on the valve/fitting assembly
Note that there are no replacement parts on the valve/fitting assembly. If the
valve fails or the fitting cannot be tightened, replace the assembly.
Leak testing the tri-column assembly
Tools Required
The following tools (not supplied in the SVE Kit) are required to perform the
leak check procedure of the tri-column assembly.
• 1/16-inch (SS) union (part no. 0100-0124)
• Vespel/graphite ferrules:
• For 250 µm column (part no. 5062-3508)
• For 320 µm column (part no. 5062-3506)
• 1/8-inch Swagelok nut (part no. 5180-4103)
• 1/8-inch solid (no hole) vespel ferrule (part no. 0100-1372) or
1/8-inch Swagelok cap
• Electronic leak detector (optional)