360 Auxiliary EPC
Theory of operation
2 of 16
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
19234-60660
(blue dot)
19231-60770
(red dot)
19231-60610
(brown dot)
10
0
80
20
30
40
50
60
70
Minimum
source
pressure
(psig)
0.1 100.0 1000.01.0 10.0
Gas flow in mL/min
Pressure requirements for AUX EPC flow restrictors
with air, nitrogen, or helium
(ambient conditions: 25°C, 14.7 psia)