310 Flame Ionization Detector (FID)
Diagnostics
22 of 24
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Diagnostics
FID ignition problems
If your FID detector is not igniting properly, check the following:
The gases are plumbed correctly.
The system is leak free.
The flow rates are set correctly.
External lines have been well purged.
The detector is contaminant free.
The ignitor is operational.
The jet is not damaged (check for scores, galls, a crimped tube) or plugged.