Inlets
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed Inlet 220
Replacement procedures
5. Remove the plumbing block from the front of the manifold (one Torx T-10
screw).
6. Remove the long screw (Torx T-20) from the top of the manifold and slide
the manifold out of the back of the GC.
7. Reinstallation is the reverse of removal.
Replacing the supply fitting on a Type 1 EPC manifold
The same replacement procedure is used to replace the supply fitting on any
of the inlet flow inlet manifolds used in the 6890 GC. See Split/Splitless Inlet
for the detailed procedure.