Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms 610
FPD checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 5]
>>Set DATE and TIME<<
6890 Analog output
Noise ([N1])
Measured at [6890 Range 5],
[Attn. 0, Chart speed 1, zero 50] ~30 mm for 1 min. measurement
Signal
6890 Range 5
3396
Zero 10 Pk wd 0.04
Att 2^ 9 Thresh 3
Cht sp 1 Stop time 11.0
Ar Rej 100000
Area, Sulfur mode 7,000,000
Agilent ChemStation
[6890 GPIB output]
Noise ([N3])
Measured by ChemStation, ASTM [N3] 5 display units (150 pA/display
units)
Signal
Area, Sulfur mode 26,000