Agilent 6890 Gas Chromatograph Service Manual
Split/Splitless Inlet 210
Leak testing—EPC and manual inlets
Performing the leak test—EPC inlets
Note Be sure to complete all of the preparation steps on pages 20 and 21 before
continuing.
1. Set the inlet to “Split Mode.”
2. Configure the column as 0 length.
Press
[Mode/Type]
Press
[Config] [Column 1] or [Config] [Column 2] and enter “0” in the first
column of the “Dim” field.