5 of 26
Jun 2001
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms 610
FID checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 8]
>>Set DATE and TIME<<
Zero 10 Pk wd 0.04
Att 2^ 4 Thresh 3
Cht sp 1 Stop time 8.75
SS/COC/PTV PPIP
Sensitivity C
14
, C
15
, C
16
area counts 125,000 125,000
Discrimination C
14
/C
16
area ratio 1.00±0.05 1.00±0.10
3396B or 3396C or 3397A integrator sample conditions with INET input
[6890 INET output, Range 8, 3396 Range 5]
>>Set DATE and TIME<<
Zero 10 Pk wd 0.04
Att 2^ 4 Thresh 3
Cht sp 1 Stop time 8.75
SS/COC/PTV PPIP
Sensitivity C
14
, C
15
, C
16
area counts 1,250,000 1,250,000
Discrimination C
14
/C
16
area ratio 1.00±0.05 1.00±0.10
Agilent ChemStation
SS/COC/PTV PPIP
Sensitivity C
14
, C
15
, C
16
area counts 4,000 4,000
Discrimination C
14
/C
16
area ratio 1.00±0.05 1.00±0.10
Agilent ChemstationASTM noise
Measured with 6890 isothermal @ 100°C
10 minute blank run, noise range(s) >1 minute
Performance + noise report
FID ASTM noise .0382 pA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals