Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Checkout Chromatograms 610
ECD checkout conditions and chromatogram
Typical values
3396B or 3396C or 3397A integrator sample conditions with analog input
[6890 Analog output, Range 0]
>>Set DATE and TIME<<
Noise ([N1])
Measured at Attn. 0, Chart speed 1 [N1] <9 mm for 1 min. measurement
Signal
Zero 10 Pk wd 0.04
Att 2^ 7 Thresh 7
Cht sp 1 [Stop time 11.0]
Ar Rej
Area, Lindane peak 84150× [N1]
3396B or 3396C or 3397A integrator sample conditions with INET input
[6890 INET output, Range 0]
>>Set DATE and TIME<<
Noise ([N2])
Measured at Attn. (–5), Chart speed 1 [N2] 11.25 mm for 1 min.
measurement
Signal
Zero 10 Pk wd 0.04
Att 2^ 7 Thresh 7
Cht sp 1 [Stop time 12.0]
Ar Rej
Area, Lindane peak 2627 × [N2]
Agilent ChemStation
Noise ([N3])
Measured by ChemStation, ASTM noise[N3] <0.05733 display units
(5 Hz/display unit)
Signal
Area, Lindane peak 1650 × [N3]