320 Nitrogen Phosphorus Detector (NPD)
Recommended settings
6 of 32
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Table 320-1 Jets and Collectors for the Capillary-Optimized NPD
The adaptable NPD fits packed columns and can be adapted to fit capillary columns.
It is shipped with the capillary column jet and standard collector. You must
change the jet to use packed columns. Instructions appear later in this
chapter.
Table 320-2 Jets and Collectors for the Adaptable NPD
The extended jets, used with the small id collectors, reduce the exposure of
the sample to heated metal and reduce tailing of some very polar components.
Type Part no. id Use with
Standard jet G1531-80560 0.29 mm Either collector
Extended jet
(optional)
G1534-80580 Either collector
Standard collector G1534-20530 7 mm
Small id collector
(optional)
G1534-20660 5 mm
Type Part no. id Use with
Capillary column jet 19244-80560 0.29 mm Either collector
Extended jet G1534-80590 Either collector
Standard collector G1534-20530 7 mm
Small id collector G1534-20660 5 mm