830 Warning Codes
Warning 119—F det calib deleted
6 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning 119—F det calib deleted
Warning 120—B det calib deleted
Status message: F det calib deleted or B det calib deleted
If you were entering a new calibration for the front or the back detector and
the 6890 GC crashed or encountered a power failure before the recalibration
was complete, this warning will appear when the instrument is turned on
again.
❐ The calibration for the module will return to the default.
Warning 121—P aux calib deleted
Status message: P aux calib deleted
If you were entering a new calibration for an aux module and the 6890 GC
crashed or encountered a power failure before the recalibration was complete,
this warning will appear when the instrument is turned on again.
❐ The calibration for the module will return to the default.
Warning 122—Host communication: data overrun
Status message: Comm data overrun
The 6890 GC received data from a computer or workstation faster than it
could process it. The data may be lost
❐ Lower the baud rate or turn on handshake mode. You can adjust both of
these parameters from the Communications control table, which is
accessed from the Options control table.
Warning 123—Host communication: data error
Status message: Comm data error
T he 68 90 GC re ceive d ba d da ta from a PC or wor ks t at io n. The da ta may b e lo st.
❐ The data bits or parity may be set incorrectly. You can adjust both of these
parameters from the Communications control table, which is accessed
from the Options control table.