Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes 830
Warning 136Front detector title mismatch
Warning 136Front detector title mismatch
Warning 137Back detector title mismatch
The title for the detector does not match the title stored in its pneumatics
module. No action is taken. If the detector is a different type from that stored
in the pneumatics module, a hardware error will be declared.
Warning 138Front injector/inlet mismatch
Warning 139Back injector/inlet mismatch
The injector is not compatible with the inlet it is intended to use.
Warning 140Front inlet Saver time changed
Warning 141Back inlet Saver time changed
While using a Volatiles Interface inlet, the Sample End time is set less than
the inlet purge time. The purge time has been changed to the Sample End time
+ 1 to avoid the conflict.
Warning 142Front inlet Purge time changed
Warning 143Back inlet Purge time changed
While using a Volatiles Interface inlet, the Sample End time is set greater than
the inlet gas saver time. The gas saver time has been changed to the Sample
End time + 1 to avoid the conflict.
Warning 144OEM Config error(s)
One or both detectors installed are not the same as the one(s) in the method.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals