850 Faults
Fault 247Front inlet module: obsolete EEPROM
16 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Fault 247Front inlet module: obsolete EEPROM
Fault 248Back inlet module: obsolete EEPROM
Status message: F inlet module rev or B inlet module rev
The front or back inlets EEPROM is not recognized by the 6890 GC. The
module is unusable.
The version of the module is not compatible with the version of the
6890 GC.
Fault 249Pres aux module: obsolete EEPROM
Status message: Aux module rev
The aux pressures EEPROM is not recognized by the 6890 GC. The module
is unusable.
The version of the module is not compatible with the version of the
6890 GC.
Fault 250Front det: non-det module
Fault 251Back det: non-det module
Status message: F det wrong module or B det wrong module
The module installed in an inlet slot is unusable. Two things could cause this
condition:
The module is installed in the wrong position. For example, a detector
module might be installed in the inlet position.
If the correct module is installed, it may be malfunctioning or its EEPROM
may be unrecognized by the 6890 GC.
Make sure that the correct type of module is installed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals