850 Faults
Fault 270Front inlet: invalid module checksum
20 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Fault 270Front inlet: invalid module checksum
Fault 271Back inlet: invalid module checksum
Status message: F inlet bad cksum or B inlet bad cksum
The inlet modules calibration is no longer valid.
Fault 272Front detector: invalid module checksum
Fault 273Back detector: invalid module checksum
Status message: F det bad cksum or B det bad cksum
The detector modules calibration is no longer valid.
Fault 274Pneu aux module: invalid module checksum
Status message: Pneu aux bad cksum
The aux pneumatics modules calibration is no longer valid.
Fault 275Front inlet: invalid constants from factory
Fault 276Back inlet: invalid constants from factory
Status message: F inlet bad fact cal or B inlet bad fact cal
The inlet modules calibration is invalid.
Fault 277Front detector: invalid constants from factory
Fault 278Back detector: invalid constants from factory
Status message: F det bad fact cal or B det bad fact cal
The detector modules calibration is invalid.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals