1150 Troubleshooting
Chromatographic symptoms
2 of 6
Jun 2001
Valves
Agilent 6890 Gas Chromatograph Service Manual
Loss of sensitivity or excessive drift
Several possible causes exist for overall deterioration of the chromatogram.
Contamination in the valve requires a thorough cleaning.
Internal leakage requires a complete disassembly and inspection of the
mating surfaces.
Poor temperature control may require a full check of electronic and
thermal components.
Lack of proper conditioning techniques, columns, etc.
Failure or deterioration of other components (columns, detectors, etc.).
Loss of peaks in specific areas of the chromatogram
Entire sections of chromatographic data can be lost due to a valve that does
not rotate or one that rotates improperly. Other than obvious component
failures (solenoid, actuator, etc.), improper adjustments and misalignments
cause most problems.
Check that adequate air (about 482 kPa or 70 psi) is supplied.
Check the valve. Is it rotating?
If the valve rotates, check for proper alignment of the actuator,
mechanical binding or slippage of connecting parts.
Check for blocked flow paths with valve in both positions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Plus and is the answer not in the manual?

Agilent Technologies 6890 Plus Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals