2-45
Making Mixer Measurements (Option 089 Only)
Isolation Example Measurements
Isolation Example Measurements
Isolation is the measure of signal leakage in a mixer. Feedthrough is specifically the
forward signal leakage to the IF port. High isolation means that the amount of leakage or
feedthrough between the mixer’s ports is very small. Isolation measurements do not use
the frequency offset mode. Figure 2-32 illustrates the signal flow in a mixer.
Figure 2-32 Signal Flow in a Mixer Example
LO to RF Isolation
LO to RF isolation is the amount the LO power is attenuated when it appears directly at
the RF port.
1. Initialize the analyzer by pressing .
2. To select the analyzer frequency range and source power, press:
( , 8722ES)
This source stimulates the mixer’s LO port.
3. To select a ratio B/R measurement, press:
Preset
Start 10 M/µ
Stop 3000 M/µ
Power 0
PWR RANGE MAN
x1
10 x1
Meas
INPUT PORTS
B/R

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals