1-55
Making Measurements
Measuring Amplifiers
Figure 1-44 Gain Compression Using Linear Sweep and
12.If was selected, recouple the channel stimulus by pressing:
13.To place the marker exactly on a measurement point, press:
14.To set the CW frequency before going into the power sweep mode, press:
15.Press .
If interpolation is on (the default setting), the calibration will be applied to the power
sweep.
16.Enter the start and stop power levels for the sweep.
Now channel 1 is displaying a gain compression curve. (Do not pay attention to channel
2 at this time.)
17.Press .
18.If was selected, press .
19.Press .
Now channel 2 displays absolute output power (in dBm) as a function of power input.
20.Press to change the scale of channel 2 to 10 dB per
division.
21.Press to change the scale of channel 1 to 1 dB per division.
D2/D1 to D2 ON
COUPLED CH OFF
Sweep Setup
COUPLED CH ON
Marker Fctn
MARKER MODE MENU
MARKERS:DISCRETE
Seq
SPECIAL FUNCTIONS
MARKER
CW
Sweep Setup
SWEEP TYPE MENU
POWER SWEEP
Chan 2 Display
DUAL | QUAD SETUP
DUAL CHANNEL ON
D2/D1 to D2 ON
MORE
D2/D1 to D2 OFF
Meas
INPUT PORTS
B
Scale Ref
SCALE/DIV
10
x1
Chan 1 1 x1

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals