EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8720E Series

Agilent Technologies 8720E Series User Manual

Agilent Technologies 8720E Series
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #398 background imageLoading...
Page #398 background image
7-28
Operating Concepts
Analyzer Display Formats
Figure 7-11 Linear Magnitude Format
SWR Format
The softkey reformats a reflection measurement into its equivalent SWR (standing
wave ratio) value. See Figure 7-12. SWR is equivalent to (1 + ρ)/(1 ρ), where ρ is the
reflection coefficient. Note that the results are valid only for reflection measurements. If
the SWR format is used for measurements of S
21
or S
12
the results are not valid.
Figure 7-12 Typical SWR Display
SWR

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals