Chapter 7 239
Digital Signal Interface Module
Operating the N5102A Module in Output Mode
From this softkey menu, set all of the clock parameters that synchronize the clocks between the N5102A
module and the ESG. You can also change the clock signal phase so the clock occurs during the valid
portion of the data. Figure 7-14 shows the clock setup menu.
Figure 7-14 Clock Setup Softkey Menu for a Parallel Port Configuration
The top graphic on the display shows the current clock source that provides the output clock signal at the
Clock Out and Device Interface connectors. The graphic changes to reflect the clock source selection
discussed later in this procedure. The bottom graphic shows the clock position relative to the data. The
displayed clock signal will change to reflect the following:
• clocks per sample selection
• clock phase choice
• clock skew adjustment
• clock polarity selection
If the device or external clock does not match the frequency, one of the following error messages will
appear on the ESG:
805 Digital module output FIFO overflow error; There are more samples being produced than can be
consumed at the current clock rate. Verify that the digital module clock is set up properly.
Inactive for a serial port configuration and the IF signal type
Active for only the Internal clock source selection
Inactive for clock rates below 25 MHz
Inactive for clock rates below
10 MHz and above 200 MHz