2 Features and Functions
Additional Features
48 U1731C/U1732C/U1733C User’s Guide
Figure 2-20 Open calibration and short calibration prompts
1 Press and hold the key for more than 1 second to
enter the open/short CAL mode for the selected frequency
and range.
2 Open/Short CAL prompts will be shown on the display.
Follow the prompts for open connector (
OPn) or short
connector (SHor) connection and press the key.
The CAL annunciator on the upper right of the display
will flash indicating that the correction is in process.
3 After the open/short CAL is completed, the LCR meter
will be restored to the normal display and ready for
normal usage.
Cal
Null
Cal
Null