ALL-TEST PRO 7™ User Manual
©2019 ALL-TEST Pro, LLC All Rights Reserved 62 Rev 06/20/2019
SPECIFICATIONS
50, 100, 200, 400, 800 Hz
Test Value Static / Reference Value Static
0.01 – 10,000 ±1%, (Dimensionless calculated value)
Repeatability ±1%, (of measured data and calculated deviations)
Repeatability ±2%, (of measured data and calculated deviations)
0.01 – 999 Ω total measurement range
0.01 – 99.9 Ω ±1%, Max Resolution: 0.01 mΩ or one count (whichever is greater)
100 Ω - 500 Ω ±1.5%, displayed as whole numbers
501 Ω - 999 Ω ±2.5%, displayed as whole numbers
Relative Accuracy “Phase to Phase” ± 0.1%
True 4-wire Kelvin measurement. (Compensation for thermoelectric offset voltages)
0.1-999 Ω ±2%, Max Resolution: 0.01 Ω or one count (whichever is greater)
Relative Accuracy “Phase to Phase” ± 0.2%
1 – 999mH total measurement range
1 – 200mH ±2%, all other values ±3%
Relative Accuracy “Phase to Phase” ± 0.2%
1– 90° ±1°, Max Resolution: 0.1°
Relative Accuracy “Phase to Phase” ± 0.1°