This test will change the output voltage in the following sequence: 
  108V for 1 minute. 
  118V for 1 minute. 
  115V for 1 minute. 
The  key (backspace) will terminate the test at any time. 
FREQUENCY 
This test will change the output frequency in the following sequence:  
  393Hz for 1 minute. 
  407Hz for 1 minute. 
  400Hz for 1 minute.  
The  key (backspace) will terminate the test at any time. 
VOLT UNBALANCE 
This test will change the output voltage for the selected phase only in the following sequence: 
  112V for 1 minute. 
  118V for 1 minute. 
  115V for 1 minute. 
The test will be repeated on three phase systems to include all three phases if the coupling is set 
to all. 
The   key (backspace) will terminate the test at any time. 
PHASE DIFFERENCE 
This test applies to three phase systems only.  The phase angle for the selected phase will 
change relative to phase A in the following sequence: 
If phase B is selected: 
 236 for 1 minute. 
 244 for 1 minute. 
 240 for 1 minute. 
If phase C is selected: 
 116 for 1 minute. 
 124 for 1 minute. 
 120 for 1 minute 
VOLTAGE MODULATION 
This test will vary the output voltage by 2.5V rms over a period of one second.  The test will last 
for 2 minutes. The  key (backspace) will terminate the test at any time. 
FREQUENCY MODULATION 
This test will vary the output frequency by 4Hz over a period of one minute.  The test will last 
for 4 minutes. The  key (backspace) will terminate the test at any time. 