Caution…
About the use:
Do not use the instrument if the internal fan is out of
order.
Before cleaning the calibrator, you must switch it off,
allow it to cool down and remove all cables.
About the liquid bath (RTC-158/250 A/B/C only):
Be careful not to overfill the well with oil.
Avoid getting silicone oil on the clothes. It is impossible
to wash off.
The oil level rises several centimetres when the
temperature is rising. Please read instructions in
section 3.3.2 about oil level. To stop overflow switch off
the main power and the oil level will decrease when
cooled down.
Carefully wipe off all silicone oil from the sensor under
test to avoid spreading of the silicone oil.
Be careful to select the right oil for the right task. Using
other than the recommended oils might cause damage
to the calibrator or degrade the performance.
Remove excess hot fluid with the outmost care, as it
might be very hot.
Do not attempt to remove hot fluid with the liquid
drainage tube, as it might melt.
About the well, insertion tube and sensor:
The well and the insertion tube must be clean and dry
before use.
Do not pour any form of liquids into the well. It might
damage the well or cause a hazard.
Do not use any alkali, acid or ionic fluids in the
aluminium well as it might be damaged.
Scratches and other damage to the insertion tubes
should be avoided by storing the insertion tubes
carefully when not in use.