127139 13 2014-07-30 15
The insertion tube must never be forced into the well.
The well could be damaged as a result, and the
insertion tube may get stuck.
Before using new insertion tubes for the calibration, the
insertion tubes must be heated up to maximum
temperature – 250°C (482°F) / 700°C (1292°F) - for a
period of minimum 30 minutes (RTC-250/700 A/B/C
only).
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on the insertion tube inside the instrument. There is a
risk that the insertion tube may get stuck if this is
allowed to happen.
If the calibrator is to be transported, the insertion tube
must be removed from the well to avoid damage to the
instrument.
The tip of the sensor should rest at the bottom of the
sensor basket for optimum results (liquid baths only).
Be careful not to submerge the handle or wire inlet of
the sensor-under-test in the fluid, as this might damage
the sensor (liquid baths only).
Note…
The product liability only applies if the instrument is
subject to a manufacturing defect. This liability becomes
void if the user fails to follow the instructions set out in this
manual or uses unauthorized spare parts.