127139 13 2014-07-30 59
The Calibration Running step 1 of 2 is started and the
temperature is heading towards step 1.
The following screen is displayed :
When the temperature has reached the stable criteria, the
calibration data will be stored and the temperature goes
towards the next set temperature.
If the workorder contains manual reading during calibration,
you will be asked to enter the Sensor Under Test temperature
before that.
The following screen is displayed :
If manual readings are specified these will have to be entered
before next step starts.
Note…
The calibration can be stopped at any time by activating “Stop
2
”, but this will erase the calibration results.
°C