ID C. T. Text Long text Action
W169
W TP Cannot open serial port for
output!
Check serial port connection.
W158
W TP Cannot open file for output! Check the output port and
presence of the output stor-
age.
W156
W TP Cannot connect to server
for output.
Check output server settings.
W139
W TP Previous “strip pads” set-
tings lost. Press “OK” (ap-
ply) before strip change.
Press Apply button to save
changes, otherwise the spe-
cial strip settings (pad order,
sediment rec., etc.) won’t be
saved.
W138
W P Server IP address or mask
format not right. (ex.:
192.168.1.12:4130)
Check and correct server IP
address or mask input.
W137
W P IP address or subnet
mask format is not cor-
rect. (i.e. 192.168.1.5/24 or
192.168.1.5/255.255.255.0)
Check and correct the analyz-
er’s IP address or mask input.
W136
W P IP address format is not
correct. (i.e. 192.168.1.12)
Check and correct the ana-
lyzer’s IP address.
W135
W TP Cannot export log, because
USB drive does not exists.
Please insert it.
Ensure the USB pendrive is
connected properly and rec-
ognized by the system.
If required, re-initialize the
USB port by pressing the
Analyticon logo on the top
right corner.
W134
W P Worklist DB failure: pos-
sible data loss! Trying to
repair. May take some
minutes, please wait.
Database failure.
The system is trying to repair
it, action is in progress. May
take some minutes, please
wait.
W134
W P Worklist DB failure: possi-
ble data loss!
Possible data loss, check
worklist. If problem occurs
multiple times, Contact your
service representative.
W133
W P Config DB failure: possible
data loss! Trying to repair.
May take some minutes,
please wait.
Data loss probably occurred.
System is trying to repair itself.
W133
W P Config DB failure: possible
data loss!
Possible configuration loss,
check database. If problem
occurs multiple times, Contact
your service representative.