Troubleshooting
ID C. T. Text Long text Action
E196
E P DB failure: configuration is
corrupted. Please check
the configuration settings.
Contact your service repre-
sentative.
EitherseePage61for‘full
database clear’.
E195
E P Worklist DB failure: cannot
write new item.
Contact your service repre-
sentative.
EitherseePage61for‘full
database clear’.
E194
E P Worklist DB failure: cannot
insert or modify item.
Contact your service repre-
sentative.
EitherseePage61for‘full
database clear’.
E193
E P Worklist DB failure: cannot
delete item.
Contact your service repre-
sentative.
EitherseePage61for‘full
database clear’.
E171
E TP Cannot export log. Ensure the USB pendrive
is connected properly and
recognized by the system. If
required, reinitialize the USB
port by pressing the Ana-
lyticon logo on the top right
corner.
E170
E TP Sample ID already exists,
please change it.
Verify and repeat the input or
use another Sample ID!
E167
E TP Operator ID already exists,
please change it.
Enter another Operator ID.
E166
E TP Password check failed,
please try again.
Enter the valid password
E165
E TP Password is too short,
please try again!
(minimum length is 3 char-
acters)
Enter a new password min. 3
characters long.
E164
E TP Password does not match,
please try again.
Re-enter password.
E163
E TP Operator does not exist,
please try again.
Enter another Operator ID.
E162
E TP Password check failed,
please try again.
Enter the valid password
E161
E TP Sample ID required.
Please set it.
Enter Sample ID
E160
E TP LOT Code required.
Please set it.
Enter LOT Code