Troubleshooting
75
ID C. T. Text Long text Action
I101
I TP Sample ID was not found,
please try again or cancel
the search.
N/A
13.1.1 Testing / Measurement Result Errors
These error codes are stored together with the results in the database perma-
nently and are also displayed after the testing procedure.
ID C. T. Long text Testing: Error Source & Action
E299
E R Head HW error: some
LEDs may be defective.
Please call Service.
Head hardware error.
Contact your service representative.
E298
E R Head HW error: voltage
out of range. Please call
Service.
Head hardware error.
Contact your service representative.
E297
E R Head HW error: software
check failed. Please call
Service.
Head hardware error.
Contact your service representative.
E296
E R Head communication
failed. Please restart the
system.
Communication with the head failed after
the measurement.
Restart analyzer and repeat the test with a
new test strip.
If the error remains, Contact your service
representative.
E282
E R Database error. Stored
item is corrupted. Please
delete item from data-
base.
Corrupted data.
Restart analyzer and repeat the test with a
new test strip.
If the error remains, Contact your service
representative.
E281
E R Database error. Missing
strip configuration data.
Please delete item from
database.
Corrupted data.
Restart analyzer and repeat the test with a
new test strip.
If the error remains, Contact your service
representative.
E280
E R Configuration error.
System configuration (or
database) failed.
Corrupted data.
Restart analyzer and repeat the test with a
new test strip.
If the error remains, Contact your service
representative.