5-19 LTE Signal Analyzer Verification, Options 541 and 542 Option Verification
5-104 PN: 10580-00255 Rev. J MT8212E and MT8213E MM
11. Adjust the MG3700A level setting to approximately –33 dBm so that the power meter
reads –50.0 dBm ± 0.5 dB.
12. Record the MT821xE Frequency Error reading into the 750 MHz, –50 dBm section of Table A-92.
13. Verify that the value is within specification.
14. Set the calibration factor frequency of the power sensor to 2150 MHz.
15. Set the MG3700A frequency to 2150 MHz. Press the Set key.
16. Adjust the MG3700A level setting with the knob so that the power meter reads –20.0 dBm ± 0.5 dB.
17. Change the MT821xE center frequency to 2150 MHz.
18. Record the MT821xE Frequency Error reading into the 2150 MHz, –20 dBm section of Table A-92.
19. Verify that the value is within specification.
20. Adjust the MG3700A level setting to approximately –33 dBm so that the power meter
reads –50.0 dBm ± 0.5 dB.
21. Record the MT821xE Frequency Error reading in the 2150 MHz, –50 dBm section of Table A-92.
22. Verify that the value is within specification.
23. For units with 20 MHz IF BW Available, which can be seen within the System Status window, repeat
Step 3 through Step 22 using the 20 MHz pattern, E-TM_3-1_20M.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2712E and is the answer not in the manual?

Anritsu Spectrum Master MS2712E Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2712E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals