Chapter 3 — T-Series Tracking Generator 3-4 Transmission Measurements
Tracking Generator MG PN: 10580-00339 Rev. C 3-5
After Normalization:
• Units in the Amplitude scale change from absolute (dBm) to relative (dB).
• The trace is centered in the graticule at 0 dB when Reference Position is set 5 (default)
and Reference Amplitude is 0.0 dB (default).
• (Trans)
is added after the Traces header on the left of the display.
After the normalization sweep is completed, reconnect the DUT between the RF OUT and
RF In connectors.
Figure 3-5 shows a filter measurement after the normalization process. Displaying the trace
required changing the Reference Position to 9 and the Scale (db range between grids).
Refer to “Transmission Measurement Menu” on page 3-11 for additional information
including Transmission Statistics and Transmission Offset.
Note
Set the appropriate input attenuation needed BEFORE turning Normalization ON.
Changes to the Amplitude menus or the Frequency menus will automatically turn
OFF Normalization.
Figure 3-5. Normalized Filter Trace