EasyManua.ls Logo

API SmartScan 350 - Figure 6: Probe Z Offset Example; Figure 7:Open Probe Settings Window; Figure 8:Probe Tip Offset Calibration Window for Probe Z Offset

Default Icon
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
offset calibration is the procedure to give the slight variation to the software. An example of
probe Z offset is shown in Figure 6.
Figure 6: Probe Z offset example.
a. Double click on "Probe Settings" under project name on left pane, and select "Z wizard" after
the "Probe Settings" window pops up.
Figure 7:Open "Probe Settings" window.
Figure 8:"Probe Tip Offset Calibration" window for probe Z offset.

Table of Contents