EasyManua.ls Logo

Applied Biosystems QuantStudio 7 Flex - Page 57

Applied Biosystems QuantStudio 7 Flex
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
b.
Conrm that the dye plate or array card matches the dye selected in the
software, then load it into the plate holder.
Load the plate or array card so that:
Well A1 of the plate or array card is in the top-left corner of the plate
adapter.
The barcode faces the front of the instrument.
Note: The name of the dy
e contained by the plate or array card is next to
the barcode on the front of it.
IMPORTANT! Plates or array cards should be loaded and unloaded by
operators who hav
e been warned of the moving parts hazard and have been
adequately trained.
c.
From the instrument touchscreen, tap to close the instrument tray.
7.
After loading the plate or array card, start the calibration:
a.
In the Dye Calibration screen, select Check the box when the dye
calibration plate has been loaded, then click Next.
b.
In the Run screen, click Start Run.
IMPORTANT! Do not aempt to open the access door during the run. The door
is locked while the instrument is in operation.
Note: Before starting the calibration, the instrument may pause (up to
10 minutes) to allow the heated cover to reach the required temperature.
Chapter 2 Calibrate the instrument
Dye calibration
2
QuantStudio
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
57

Table of Contents

Related product manuals