Chapter 3 Install the Colocated Layout
Perform the RNase P Experiment
41
Applied Biosystems StepOne
™
and StepOnePlus
™
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
Notes
Run the Experiment
After preparing the TaqMan
®
RNase P Fast Instrument Verification Plate, load the plate
into the instrument and start the run.
Load the
Instrument
PHYSICAL INJURY HAZARD. During instrument operation, the
temperature of the sample block(s) can exceed 100 ° C. If the instrument has been used
recently, keep your hands away until the sample block(s) reach room temperature.
IMPORTANT! Wear powder-free gloves when you handle the
RNase P plate
.
1. Open the instrument drawer.
2. Place the RNase P plate in the sample block(s) so that the A1 position is at the back-
left corner of the sample block(s).
3. Close the instrument drawer carefully.
A1 A1
StepOnePlus
System StepOne
System