SECTION 2 INSTALLATION AND PREPARATION FOR USE
______________________________________________________________________
2-2
Press ARROW AHEAD key Verify REVIEW MODE screen displays.
Press MENU/EXIT key Verify REVIEW MODE screen exits.
Press MENU/EXIT key Verify Setup menu displays.
Press UP ARROW key Verify highlight bar moves up with each press.
Press DOWN ARROW key Verify highlight bar moves down with each press.
Press SELECT key Verify that highlighted menu line is selected.
Press MENU/EXIT until BIS banner is displayed.
9. End of install.
2.2 ENVIRONMENT
2.2.1 Shipping and Storage Environment
The monitor and its accessories can be stored or shipped within the following
environmental limits. Note that these limits apply to non-operational storage and
shipping situations.
Temperature -20°C to +60°C
Humidity 15% to 95% (non-condensing)
Pressure 360 mmHg to 800 mmHg
Protect the monitor from sudden temperature changes that can lead to condensation
within the instrument. To minimize condensation, avoid moving the system between
heated buildings and outside storage. Once moved inside, allow the monitor to stabilize
in the unopened shipping container at the inside ambient temperature before unpacking
and placing into service. Before operation, wipe down all visible condensation and allow
the system to reach equilibrium at room temperature.
2.2.2 Operating Environment
The A-2000 is not designed for use in areas containing flammable gases or vapors.
WARNING!
EXPLOSION HAZARD: DO NOT USE THE A-2000 IN A FLAMMABLE
ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE
ANESTHETICS MAY OCCUR.
Temperature: The Aspect A-2000 Monitor is designed to operate safely at a room
temperature of 5°C to 40°C. Conditions that exceed these limits could affect reliability.
Humidity: The monitor is designed to operate within specifications at a relative non-
condensing humidity of 15% to 95%.
Pressure: The monitor will operate satisfactorily at or above sea level, and is unaffected
by extremes or changes in altitude within atmospheric pressures of 360 mmHg to 800
mmHg.