EasyManua.ls Logo

Aspect A-2000 BIS - Figure 10 the Sensor Check (bis Quatro and bis Extend)

Aspect A-2000 BIS
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 3 OPERATING THE A-2000
______________________________________________________________________
3-4
Figure 10 The Sensor Check (BIS Quatro and BIS Extend)
The impedance value for each electrode, in kilo ohms, appears on the screen along with
its status:
PASS - An electrode passes if the impedance for that electrode is less than 7.5 kilo
ohms. The ground electrode (element #2) must be less than 30 kilo ohms to pass.
HIGH - An electrode is labeled "HIGH" if its impedance value is above 7.5 kilo ohms. As
long as the combined impedance (elements #1 and #3) is less than 15 kilo ohms, the
sensor check will be considered successful. (BIS Quatro and BIS Extend also check
combined impedance of elements #1 and #4). If the combined impedance is over the 15
kilo ohms limit, you will need to re-prep the electrodes and check all connections. The
monitor will continue to check impedance until it is acceptable.
NOISE - If the signal from the electrode goes beyond the measurable range, the label
"NOISE" will appear.
LDOFF - If a lead is too loose or disconnected, the label "LDOFF" will appear.
The monitor continues updating the values until all impedance values are acceptable. To
end the impedance test, press [MENU/EXIT]. The Sensor Check impedance test must
be successfully completed before normal processing resumes.
Note:
All signal status messages will be suppressed for 60 seconds after a
sensor check has exited. This provides the time for signal quality to
recover.
The Sensor Integrity Check (BIS XP Platform only) is also initiated each time that a
sensor is connected to the PIC. It checks to make certain that a valid, unexpired sensor
is in use.

Table of Contents