SECTION 6 PREVENTIVE MAINTENANCE, CARE AND CLEANING
______________________________________________________________________
6-4
6.4 SYSTEM CHECKOUT
1. Secure all components.
Monitor with power cable
DSC
PIC (Patient interface cable, connects DSC to patient)
BIS Sensor or test sensor.
2. Connect power cable to monitor, plug power plug into appropriate wall outlet.
3. Power up monitor by pressing power switch (right lower side of case).
Verify beep tone as power switch is activated.
Verify fan (rear case wall) moves air outward.
Verify all power on self-tests complete successfully (approx. 30 seconds).
Verify next screen says ‘CONNECT DSC CABLE TO FRONT OF MONITOR.’
4. Connect DSC with PIC and Sensor.
Verify DSC test completes.
Verify SENSOR CHECK screen displays.
5. Exit from SENSOR CHECK screen by passing impedance check or by exiting with
MENU/EXIT key.
6. Disconnect power cord from rear of monitor.
Verify ‘OPERATING ON BATTERY BACKUP (E33)’ is displayed.
Verify battery icon displays in BIS banner.
7. Reconnect power cord.
Verify battery icon is not displayed in BIS banner.
Verify ‘OPERATING ON BATTERY BACKUP (E33)’ is not displayed.
8. Do keyboard checkout:
Press SILENCE key Verify icon shows at BIS banner.
Press ARROW BACK key Verify REVIEW MODE screen displays.
Press MENU/EXIT key Verify REVIEW MODE screen exits.
Press ARROW AHEAD key Verify REVIEW MODE screen displays.
Press MENU/EXIT key Verify REVIEW MODE screen exits.
Press MENU/EXIT key Verify Setup menu displays.
Press UP ARROW key Verify highlight bar moves up with each press.
Press DOWN ARROW key Verify highlight bar moves down with each press.
Press SELECT key Verify that highlighted menu line is selected.
Press MENU/EXIT until BIS banner is displayed.
9. End of checkout.